RTD calibrator - PCE-RTD 20 - Simulation of resistors and sensors ... - PCE Instruments
RTD calibrator / Simulation and measurement of resistance, Pt sensors, current and voltage / Li-Ion battery / Continuity check / Ramp function
The RTD calibrator, model PCE-RTD 20, was designed to simulate resistors and resistance temperature sensors. This allows you to use the RTD calibrator as a calibrator of temperature measuring devices. In addition to simulating resistances, the RTD calibrator has a function for measuring voltage, current and resistance. These two functions work independently of each other. In addition, the two connections are galvanically separated.
The RTD calibrator integrates different characteristic curves for RTD sensors, which allows thermometers of different types to be correctly calibrated. The accuracy of the RTD calibrator is 0.02% for all functions and for all ranges. In addition to determining the temperature or resistance value, the PCE-RTD 20 allows you to work in Rampa mode. The user can decide to use Ramp mode continuously or at set intervals.
The RTD calibrator is equipped with a mini USB port through which you can charge the battery and which is also used to transfer measured values.
- Simulation and measurement mode
- Battery
- Graphic screen
- 24 V power supply for loop
- HART compatible
- Manual mode and ramp function
- Continuity test
- Data logging function
DELIVERY CONTENT
1 x RTD PCE-RTD 20 calibrator
3 x Cables with 2 mm connector
3 x 2mm cables with alligator clips
3 x Adapters between 2 mm and 4 mm
1 x Mini USB Data Cable
1 x 5V/1A Power Source
1 x Carrying bag
1 x Software
1 x User Manual
SPECIFICATIONS
Setting |
Range |
Resolution |
Precision |
DC voltage V |
0 ... 30 V |
0.001V |
±0.02 % of the value + 2 d. |
CC mA
|
0 ... 24 mA
|
0.001 mA
|
±0.02 % of the value + 2 d.
|
Resistance
|
0 ... 400 Ω 400 ... 4000 Ω*
|
0.01Ω 0.1Ω
|
±0.02% of value + 0.01Ω ±0.02 % of the value + 0.015Ω
|
Pt10 ... Pt1000
|
-200 ... 200 °C 200 ... 600 °C 600 ...850 °C
|
Pt10... Pt 400: 0.01 °C Pt500 ... Pt1000: 0.1 °C
|
±0.2 °C ±0.3 °C ±0.1 °C
|
Ni100
|
-60 ... 180 °C
|
0.01 °C
|
±0.1 °C
|
Ni120
|
-80 ... 260 °C
|
0.01 °C
|
±0.1 °C
|
Cu10
|
-200 ... 260 °C
|
0.01 °C
|
±0.2 °C
|
* In the measurement mode of 4 cables, a resolution of 0.01 Ω is possible, on one range from 0 ... 1600 Ω. The indicated accuracy is applicable to the 4-cable measurement mode. Measuring 3 cables increases the inaccuracy by 1ºC (Pt10/Cu10), 0.6ºC (Pt50/Cu50) and 0.4ºC (for the rest of the models). |
|
Simulation parameters |
Range simulation |
Resolution |
Accuracy* |
Resistance |
0 ... 400 Ω 400 ... 4000 Ω |
0.01Ω 0.1Ω |
±0.02% of value + 0.01Ω ±0.02 % of the value + 0.015Ω |
Pt10 ... Pt1000
|
-200 ... 200 °C 200 ... 600 °C 600 ... 850 °C
|
Pt10... Pt 400: 0.01 °C Pt500 ... Pt1000: 0.1 °C
|
±0,15 °C ±0.25 °C ±0,15 °C
|
Ni100
|
-60 ... 180 °C
|
0.01 °C
|
±0,15 °C
|
Ni120
|
-80 ... 260 °C
|
±0,15 °C
|
0.01 °C
|
Cu10
|
-200 ... 260 °C
|
0.01 °C
|
±0.8 °C
|
* Accuracy valid for a current of ✱0.2 mA respectively ✱0.4 mA. |
Compatible RTD sensor
|
Pt10 (385), Pt50 (385), Pt100 (385), Pt200 (385), Pt400 (385), Pt500 (385), Pt1000 (385), Pt10 (3926) Ni100 (672), Ni (618), Ni120 (672), Cu10 (427) Cu50 (427), Cu100 (427)
|
|
Generic specifications |
Screen modes |
Measure: mA / V / Ω / RTD Simulation: Ω / RTD |
Temperature units |
°C / °F / K |
RTD measuring current |
Approximately 300 µA |
Maximum current simulation |
3 mA (0 ... 650 Ω) I |
Maximum input voltage |
30 VDC |
Temperature coefficient |
|
Entry impedance |
Voltage measurement: >1 MΩ Current measurement: 10 Ω |
Response time |
|
Screen refresh rate |
10Hz |
Data logging |
Internal memory, 150,000 values |
Interface |
USB2.0 |
Screen |
2.4" TFT LCD / 240 x 320 pixels / Backlight |
Loop output voltage |
24 V DC / 24 mA |
HART mA loop resistance |
250 Ω ±20 % |
Special functions |
Passage and ramp function Manual and automatic mode √x, x2: for the measurement function |
Continuity test |
Limit value control up to 100 Ω |
Food |
3.7 V / 2300 mAh Li-Ion battery |
Load duration |
5 hours |
Power supply |
Input: 100 ... 240 V AC / 50 / 60 Hz Sortie : 5 V / 1 A DC |
Battery operating time |
Approximately 3 p.m.: RTD/Ω simulation and measurement mode with dim LCD lighting Approximately 8 h: Modo measures 12 mA (24 V) with weak LCD lighting |
Dimensions |
162 x 82 x 40mm |
Weight |
Approximately 300g |
Protection category |
IP20 |
Working conditions |
Battery: 0 ... +55 °C / 30 ... 90 % H.r. Electrical network: 0 ... +45 °C / 30 ... 90 % H.r. |
Storage conditions |
-20 ... +60 °C / 30 ... 90 % H.r. without condensation |
Warm-up time |