Ammeter - PCE-CM 3 with open forks - PCE Instruments
With open forks / Measurement of several electrical parameters
The open-fork ammeter clamp is used for quick and simple AC measurements. All you have to do is insert the open forks into the one in the wire. The tweezers are specially designed to measure alternating currents in electrical enclosures and in any circuit that does not need to be interrupted. The range current measurement ranges from 0 to 200 A. In addition to measuring current, you can also measure AC and DC voltages up to 600 V, resistances, capacitance and temperature. Measurement values are shown on the backlit LCD display.
Its compact dimensions and low weight, as well as its durable housing, are some of the features of this current clamp with open forks. All this makes the ammeter an ideal tool for any installer or maintenance technician
Current measurement up to 200 A AC
Voltage measurement up to 600 V AC/DC
Multimeter function
Compact dimensions
Battery powered
Backlit LCD screen
Optional: ISO calibration certificate
DELIVERY CONTENT
1 x PCE-CM 3 open fork voltmeter
2 x test cables
2 x Measuring tips
1 x Adapter for thermo elements
1 x Thermo element type K
2 x 1.5 V batteries, type AAA
1 x User manual
SPECIFICATIONS
DC voltage
|
Range
|
Resolution
|
Precision
|
|
4 V DC
|
1mV
|
± (1.2 % of value + 2 d)
|
40 VDC
|
10mV
|
± (1.5 % of value + 2 d)
|
400 V DC
|
100 mV
|
± (1.5 % of value + 2 d)
|
600 V DC
|
1V
|
± (2 % of value + 2 d)
|
Input impedance
|
10 MΩ
|
Overvoltage protection
|
600 V DC 600 V AC RMS
|
AC voltage
|
Range
|
Resolution
|
Precision
|
|
4 V AC
|
1mV
|
± (1.5 % of value + 5 d)
|
40 V AC
|
10mV
|
± (1.5 % of value + 2 d)
|
400 VAC
|
100 mV
|
± (1.5 % of value + 2 d)
|
600 V AC
|
1V
|
± (2 % of value + 2 d)
|
Input impedance
|
10 MΩ
|
Overvoltage protection
|
600 V DC 600 V AC RMS
|
Range frequency
|
50...400Hz
|
Alternating current
|
Range
|
Resolution
|
Precision
|
|
200 A AC
|
100mA
|
± (3.0% of value + 5 d)
|
Overload protection
|
200 A AC
|
Range frequency
|
50 ... 60 Hz
|
Resistance
|
Range
|
Resolution
|
Precision
|
|
400Ω
|
0.1Ω
|
± (1% value + 4 d)
|
4kΩ
|
1Ω
|
± (1.5 % of value + 4 d)
|
40kΩ
|
10Ω
|
± (1.5 % of value + 4 d)
|
400kΩ
|
100Ω
|
± (1.5 % of value + 4 d)
|
4 MΩ
|
1kΩ
|
± (2.5% of value + 4 d)
|
40 MΩ
|
10kΩ
|
± (3.5% of value + 4 d)
|
Overvoltage protection
|
250 V DC 250 AC RMS
|
Frequency
|
Range
|
Resolution
|
Precision
|
|
40Hz
|
0.01Hz
|
± (1% value + 2 d)
|
400Hz
|
0.1Hz
|
± (1.2 % of value + 2 d)
|
4 KHz
|
1Hz
|
± (1.2% of value + 2 d)
|
40 KHz
|
10Hz
|
± (1.2 % of value + 2 d)
|
400 KHz
|
100Hz
|
± (1.2 % of value + 2 d)
|
10 MHz
|
1 KHz
|
± (1.2 % of value + 2 d)
|
Input sensitivity
|
10 mV RMS
|
Overvoltage protection
|
300 V DC 300 AC RMS
|
Work cycle
|
Range
|
Resolution
|
Precision
|
|
0,1 % ... 99,9 %
|
0.1%
|
±1.2% of value ±2 d
|
Input sensitivity
|
1 V RMS
|
Pulse width
|
> 100 µs, |
Overvoltage protection
|
300 V DC 300 AC RMS
|
Temperature
|
Range
|
Resolution
|
Precision
|
|
-20 ... 760 °C
|
0.1°C 1°C
|
± (3 % of value + 5 °C)
|
4 ... 1400 °F
|
0.1°F 1°F
|
± (3 % of value + 8 °F)
|
Overvoltage protection
|
250 V DC 250 AC RMS
|
Thermo element connection
|
Adapter with connection for type K
|
Ability
|
Range
|
Resolution
|
Precision
|
|
4 nF
|
0.1nF
|
± (5% of value + 20 d)
|
40nF
|
1nF
|
± (3% of value 5 d)
|
400nF
|
10 nF
|
± (3% of value 5 d)
|
4 μF
|
100 nF
|
± (3% of value 5 d)
|
40 µF
|
1µF
|
± (3% of value 5 d)
|
100µF
|
10µF
|
± (3% of value + 10 d)
|
Overvoltage protection
|
250 V DC 250 AC RMS
|
Function
|
Test conditions
|
Indication
|
Diode testing
|
Test current: 0.5mA Tension inverse : 1,5 V
|
Diode forward voltage
|
Continuity test
|
Open circuit voltage: 0.5 V
|
Noise when resistance |
Overvoltage protection
|
250 V DC / 250 AC RMS
|
General technical specifications
|
Opening of the clamp
|
17 mm approximately
|
Screen
|
LCD 4000 backlit digits
|
Continuity test
|
Noise when resistance is less than 50 Ω
|
Test current
|
0.5 mA approximately
|
Open circuit voltage
|
|
Battery level indication
|
Symbol on screen when the voltage is low
|
Exceeding indication
|
Yes, the display shows "OL"
|
Quota update values on screen
|
3Hz
|
Temperature sensor
|
Thermo element type K
|
Input impedance
|
10 MΩ
|
Working conditions
|
5 ... 40 °C / 80% H.r., without condensation
|
Storage conditions
|
-20 ... 60 °C / 80% H.r., without condensation
|
Maximum operating altitude
|
|
Food
|
2 x 1.5 V batteries, type AAA
|
Automatic shutdown
|
30 minutes
|
Security
|
IEC1010-1 (2001) CAT II 1000 V CAT III 600 V
|
Degree of contamination
|
2
|